Résumé

Atomic force microscopy (AFM) experiments were performed to study the behaviour of AFM cantilevers under an external magnetic field B and temperature field produced by a coil with an iron core. Four cantilever types were studied. Forces were measured for different B values and at various coil-to-cantilever separation distances. The results were analysed on the basis of a phenomenological model. This model contains the contribution of two terms, one monopole-monopole interaction at short distance, and one apparent paramagnetic interaction in del B-2 at large distance, which represents the temperature effects. We observe a good agreement between the model and the experimental data.
Fichier principal
Vignette du fichier
Kumar_15518.pdf (1.05 Mo) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)
Loading...

Dates et versions

hal-01279286 , version 1 (25-02-2016)

Identifiants

Citer

Mithlesh Kumar, Richard Arinero, Wladimir Bergez, Philippe Tordjeman. Thermo-magnetic behaviour of AFM–MFM cantilevers. Measurement Science and Technology, 2015, 26 (8), pp.085002. ⟨10.1088/0957-0233/26/8/085002⟩. ⟨hal-01279286⟩